55/5500 Executive EMS Aircraft Interiors
The 5500 Module is especially designed for larger executive-style aircraft, such as an Airbus A320 or a Boeing 727. This module is custom-built to provide comfort for the patient, even on long-distance flights and for your specific auxiliary medical equipment.
The stretcher pad is 4 inches (7.6 cms.) thick and 27 inches (68.6 cms.) wide. The front of the base has two patient transport legs which pull out from the base to aid in stabilizing the ground stretcher when loading the patient onto the 5500 Stretcher. The headrest and the legrest both rise to provide for position changes for the patient.
The overhead is most unique. While in flight, your auxiliary medical equipment (such as the Propaz, Oxylog and Infusion Pump) will be attached on the overhead ledge. The overhead typically has the following features:
- Electrical supply panel, which includes digital oxygen readouts, the coverter/inverter operating lights, switches and circuit breakers, the AC selector and DC circuit breaker, two air switches, and a florescent light with dimmer
- Three oxygen latch valves (outlets), air outlet latch valve, and oxygen and air pressure gauges
- Four IV Pole latches
The base of the 5500 has six built-in drawers. The medical equipment stowage drawer and the defibrillator drawer are equipped with 5 electrical outlets so your medical equipment can be recharging when not in use. The oxygen cylinder drawer has space for two oxygen bottles and pulls out to easily remove the cylinders when they need recharging.
| Bench Length | 79" (200.6 cm) |
|---|---|
| Width | 25" (63.5 cm) |
| Height | 25" (63.5 cm) to top of bench |
| Base Unit Weight | 378 lbs. (171.8 kg) |
| Air Pump Capacity | 11 lpm. @ 50 psi. |
| Vacuum Pump | 14 lpm. @ 14 in hg. |
| Oxygen Supply | 7,000 Liter |
- Seat Track Adapter for easy mounting into the existing aircraft seat tracks.
- Oxygen Drawer with mounts / plumbing for two ( 2 ) composite O2-cylinders, total 12.000 ltrs. O2, Oxygen Selector Valves and Connector Panel for individual O2 system selection and emergency isolation, access to the O2 shut-off valves and connectors. O2 overboard valve.
- Drawer Section with different sized internally lighted, lockable drawers for medical devices and medical supplies. Mounting of different medical devices for storage and recharge of medical device accumulators during flight.
- Overhead Supply Panel ( OSP ) contains all switches, breakers, and gauges for operating the PTU. In addition, the OSP has a large shelf with special tracking for mounting of medical equipment and monitors and hanging IV bags. OSP contains 115 or 230 volt electrical outlets, oxygen and air latch valves for connecting flow-meters and other equipment.
- Electrical Installation with integrated dual inverters ( 1000 Watt ) and air pumps. Audio/Visual Inverter out warning and emergency isolation.
- Wide-body Stretcher with fully moveable back- and footrest for improved comfort and patient care. 4" Stretcher Pad with "thermal-activated" mattress and vinyl covering for ease of cleaning. Four point harness for securing patient in transport. Mounting provisions for: additional IV-poles, examination light and medical devices.